10 . When Not Using the Scanner Head Unit
10 . When Not Using the Scanner Head Unit (MF-C2000A-
MC/SC Only)
The scanner head unit is controlled by CL-E100A built in MF-C2000A-MC/SC.
When replacing the scanner head unit with the output unit and operating the laser
equipment alone, it is necessary to separate connection between MF-C2000A-MC/
SC and CL-E100A.
Disconnection is performed by the following procedure. They can be reconnected by
reversing the procedure.
⇒
When changing availability of the scanner head, the output correction may be
required. For the output correction, contact us.
Item required
Phillips screwdriver, EXT.I/O (1) short-circuit connector, and LASER IF (2) short-
circuit connector
Operating Procedure
(1) Turn OFF the MAIN POWER switch and open the connector cover on the rear.
(2) Disconnect the cable connected to the EXT.I/O (1) connector and connect the at-
tached EXT.I/O (1) short-circuit connector.
(3) Disconnect the cable connected to the LASER IF (2) connector on the connector
panel 2 and connect the attached LASER IF (2) short-circuit connector.
EXT . I/O (1) connector
LASER IF (2) connector
Содержание MF-C2000A Series
Страница 1: ...AA04OM1209213 07 FIBER LASER WELDER MF C2000A Series OPERATION MANUAL ...
Страница 21: ...Introduction Part ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 43: ...Installation and Preparation Part ...
Страница 44: ......
Страница 73: ...Operating Part ...
Страница 74: ......
Страница 78: ......
Страница 142: ......
Страница 208: ...208 MF C2000A Series 4 Connector Functions REM I L Connector E STOP Connector Refer to the Operating Part Chapter 4 ...
Страница 212: ......
Страница 213: ...Maintenance Part ...
Страница 214: ......
Страница 232: ......
Страница 244: ......
Страница 245: ...Appendixes ...
Страница 246: ......