6.4
Sensor
This option is used to calibrate the selected sensor. We recommend calibrate the sensor before printing when
changing the media.
Item
Description
Default
Auto calibration
This option is used to set the media sensor type and calibrate the
selected sensor automatically. Printer will feed 2 to 3 gap labels to
calibrate the sensor sensitivity automatically.
N/A
Manual setup
In case “Automatic” cannot apply to the media, please use “Manual”
function to set the paper length and gap/bline size then scan the
backing/mark to calibrate the sensor sensitivity.
N/A
Threshold detect
This option is used to set sensor sensitivity in fixed or auto.
Auto
Maximum Length
This option is used to set the maximum length for label calibration.
254 mm
Advanced
This function can set the minimum paper length and maximum
gap/bline length for auto-calibrate the sensor sensitivity.
0 mm
Menu
Sensor
Auto calibration
Gap
Black mark
Continuous
Preprint
Manual setup
Gap
Black mark
Continuous
Threshold detect
Auto
Fixed
Maximum length
Advanced