Calibration
52
After calibration is finished, confirm that the readings from the CRT-300
Cellcorder are within tolerance as compared to the values in the Actual Input
column. If any reading is out of tolerance, the unit must be recalibrated. If all
readings are within specifications, proceed with the next section.
16.9 Recording the As Left K-Constants
Shift+Serv|Password|Calibration|Cal-K|
Record the after-calibration K-Constants onto the Calibration form. To
display the constants, from the Calibration menu, press F2 Cal-K. Press the
Up/Down arrows to display three screens of K-constants. Record these
values in the Calibration Form K-Constant After Cal column. When finished,
press Cancel.
16.10 Backing Up and Restoring the Calibration
Refer to the
Cellcorder Battery Analysis System User's Guide
for
information about backing up and restoring Cellcorder calibration.