Calibration
51
16.8 ICr Channel Setup and Calibration
Shift+Serv|Password|Calibration|Test
I
Cr Channel
To calibrate the ICr (intercell) channel, connect the Cellcorder resistance
leads (red lead not used) and mV calibration source power supply as shown
in the following figure.
Figure 27. Intercell Channel Setup
NOTE: If Linearity tests fail to meet min/max range, do not adjust the
Up/Down arrow keys. You must recalibrate the initial setting where you
used the arrow keys, then retest linearity.
Figure 28. Intercell Channel Menus
1)
With the millivolt (mV) Cal source at 0 volts, connect the DVM to
points E and F at the Cal source.
2)
From the Calibration menu, select ICr channel.
3)
Adjust the Cal source so the DVM indicates 21mV
1mV.
4)
Press the Up/Down arrows on the CRT until the voltage on the CRT
matches the reading on the DVM.
NOTE: The CRT displays voltage in microvolts (
V).
5)
Record the actual input and the CRT reading on the Calibration form.
6)
Press F2 Accept.
7)
From the Calibration menu, select ICr channel.
8)
Linearity: Adjust the Cal source to 0.6mV
0.05mV on the DVM and
record the actual input and CRT reading.
9)
Linearity: Adjust the Cal source to 80mV
1mV on the DVM and
record the actual input and CRT reading.
10)
Press F3 Cancel.
11)
Remove the DVM leads from points E and F.
12)
Set the Cal source to 0 volts.
Calibration
Source (mV)
Power Supply
CRT-300
Cellcorder
Red/Black Clamp
E
( + )
Black Clamp
F
( – )