Setting Up and Editing a Site
23
NOTE: When a site is deleted, the last site shifts to fill the deleted site. For
example, if site 003 of 006 is deleted, site 006 becomes 003 of 005. You may
use the BAS program or CRT-300 Previewer to sort the sites on location,
battery or string name.
8.7
Selecting a Site for Test
Before reading voltage or resistance, you should select a site. A site setup is
not required for testing, but is useful for managing many sites. If a site is
selected before testing, both the readings and site setup will be saved in
the
.CDF
file.
Main Menu > Load Data|Site Setup
To select a site, at the Main Menu, select Load Data|Site Setup. Press the
Right/Left Arrows to step to a site, then press Select.
Shift+Site
If a site has been selected and you want to take readings with no site
selected, press Clear. Refer to
Unlinking a File and Clearing Thresholds
for
more about Clear. To select a different site, press Load.
TIP
: To determine at any time what site is selected, press Shift+Site.