![AIXTRON CRIUS II Скачать руководство пользователя страница 145](http://html1.mh-extra.com/html/aixtron/crius-ii/crius-ii_system-manual_2883093145.webp)
AIXTRON
- Dokumentation
CONFIDENTIAL
145
6
cr
ius_I
I_e
n_
00,
Ed
it
io
n
06/
201
0
System Manual
Maintenance
CRIUS II
Information on elastomer seals
Fig. 6-3
Inspecting a seal
A Cuts and porous areas
B Enlargement, swelling
Cleaning
3
Pull the seal through a dry and clean cloth to remove coarse soiling
4
Pull the seal through a cloth moistened with isopropanol
Inserting
5
Insert the seal
6
Check the excess of the seal in the groove. See Fig. 6-4,
145.
•
Seal thickness 5 mm: Excess = 0.8 mm
•
Seal thickness 6 mm: Excess = 1.1 mm
Fig. 6-4
Excess of a seal in the groove
A Excess
NOTE:
Do not apply vacuum grease to seals used in the vacuum system or pro-
cess gas system of the CRIUS II!