Monnal T75 user manual
Automatic alarm thresholds
VCV
VPC
SIMV
PSIMV
PSV
PSV
NIV
CPAP
APRV
Duo
Levels
PRVC
PS PRO
P peak
(cmH2O)
cmH2O
P peak
mea-
sured +
33%
P peak
mea-
sured
+max.
(20%,
+5)
--
P peak
memax. (20%, +5)
--
P peak
memax.
(20%, +5)
Pmin
(cmH2O)
cmH2O
P peak
mea-
sured -
33%
P peak
mea-
sured -
max
(20%, 5)
--
P peak measured - max (20%,
5)
--
P peak
measured - max
(20%, 5)
P plat
(cmH2O)
cmH2O
P plat
mea-
sured
+ 5
OFF
--
OFF
OFF
VTi low
(mL)
mL
VTi
measured - 75%
OFF
VTi
measured - 75%
--
--
--
VTi
measured - 75%
VTi high
(mL)
mL
VTi
me 50%
2000
mL
VTi
me 50%
--
--
--
VTi
me 50%
MVe
low (L/
min)
L/min
MVe measured - 50%
MVe
high (L/
min)
L/min
MVe me
50%
MVe me 75%
MVe
mea-
sured +
50%
MVe
mea-
sured +
75%
RR low
(c/min)
c/min
RR measured - 50%
RR high
(c/min)
c/min
RR me 50%
VMCO2
low
mL/min VMCO2 measured – max(- 50%, 10)
VMCO2
high
mL/min VMCO2 measured – max(+ 50%, 30)
VMalv
low
L/min
VMCO2 measured – max(- 50%, 10)
high
VMalv
L/min
VMalv measured – max(+ 50%, 5)
132