Manual adjustment
The device supplies a nebulization pressure of 1.2 bar. The user must refer to the nebulizer instruc-
tions to find out the correct nebulization flow rate for this pressure, and then adjust it accordingly on
screen.
Nebulization time
This defines the duration of the nebulization session.
Periodicity
This defines the frequency of the nebulization sessions.
Press the
[Periodicity]
button to set the frequency of nebulization (hours).
If necessary, you can activate just one nebulization session which will not be repeated periodically. To
do this, set the PERIODICITY parameter to “NONE”.
To go back to the previous screen, press
[Back]
.
6.3.5 Loop curves
The following types of loop curves can be displayed:
•
Volume / Pressure;
•
Flow rate / Pressure;
•
Volume / Flow rate;
•
CO2 / Volume (if the CO software option is present).
To access the loop curves:
1.
Press the
[Curves]
key in the right-hand bar
2.
Press the
[Loops]
key.
To access these various types of loop curve, touch the curves zone until the desired curve is displayed.
To adjust the volume, pressure, flow rate, or CO2 scale, press the appropriate key (V mL, P cmH2O, D
L/min, CO2 mmHg, CO2% or CO2 kPa) until the desired scale is displayed.
To freeze the loop curve, press the
[Freeze]
key.
When the loop curve is frozen, two cursors appear and allow the value of the curve to be measured at
the point defined by the cursor.
To select the cursor, press the control wheel.
72