4 Use
The
[History]
key opens a window which displays a chronological list of the last 200 alarms.
For each alarm there is a date, priority (color), time, activation or deactivation (ON or OFF), name, and
for physiological alarms, the alarm threshold setting at the time of activation.
Use the control wheel to browse the list.
To exit the alarm history, press the control wheel.
Note:
When a complete power cut occurs (mains, external and internal batteries), the histo-
ry is preserved and can be displayed after the power supply has been restored. These data
are preserved thanks to a back-up battery with a working life of approximately two years. An
alarm is triggered when this battery runs low. The battery should then be replaced.
Note:
When the log reaches its maximum size, the oldest events will be overwritten by in-
coming events.