Monnal T75 user manual
CPAP MODE (Continuous Positive Airway Pres-
sure)
CPAP
Ventilation at two pressure levels
Duo-Levels
Pressure-regulated volume controlled ventilation PRVC
Spontaneous ventilation with inspiratory assis-
tance, PEEP and servomechanism frequency
PS-Pro
Airway Pressure Release Ventilation
APRV
Inspiratory trigger system
Main inspiratory trigger in flow rate, secondary in pressure:
Turn the control wheel to set the inspiratory trigger to between 1 and 15 L/min or between 0.2 and 3
cmH2O. A pressure threshold
varying from 0.2 to 5 cmH2O is associated with it.
At the time of a patient demand, satisfying one of the conditions (flow rate or pressure) will trigger
an inspiratory cycle.
Expiratory trigger system
During any spontaneous cycle, the switch to expiration occurs as soon as one of the following crite-
ria is encountered:
- expiratory flow trigger (Trig. E); Expiration as soon as the inspiratory flow rate reaches x% of peak
inspiratory flow rate
- expiratory pressure trigger (not adjustable): expiration as soon as an overpressure of 3 cmH2O is
detected on the inspiratory
pressure signal
- expiratory time trigger (TI max.): expiration as soon as the insufflation time reaches the maximum
Ti setting (Timax).
Automatic Test
Tests initialization
Checking the integrity of sensors to begin the
tests
Circuit rinsing
Eliminating the oxygen present in the system
Pneumatic check
Checking the integrity of the actuators of the in-
spiratory
and expiratory branches
Calibrating the oxygen and expiratory flow sen-
sors
Mixer check
Checking the mixer