14.2.1 Configurable measuring ranges
Description
Range
Exp Meas. range
Units Resolution
1. * Temperature T, t
B-01
D t
-1
-20..+70.00 °C
0.1 K
2. * Flow, v with PC (TH4)
B-02
D v
-3
0.08... 2.000
+
m/s
0.001 m/s
2. * Flow, v with PC (TH5)
B-02
D v
-2
0.2... 20.00
+
m/s
0.01 m/s
3. * Atm. pressure AP, p
B-03
D p
-1
300...1100.0 mbar 0.1 mbar
14.3 Technical data
Operative range
-20 to +70 °C
Flow
FVAD35-TH4
Measuring range
0.080 to 2 000 m/s
Accuracy
±(0.04 m/s +1% of meas. val.)
±0.5% of meas. val. / °C (0.3 to 2m/s)
FVAD35-TH5
Measuring range
0.20 to 20.00 m/s
Accuracy
±(0.2 m/s +2% of measured value)
±0.3% of measured value / °C (0.3 to 20 m/s)
Response time
<1.5 seconds
Temperature compensation
0 to +50 °C
Temperature
Measuring range
-20.0 to +70.0 °C
Accuracy
±0.7 °C at 0 to +50 °C
Response time
10 seconds
Atmospheric pressure
Measuring range
300 to 1100 mbar
Accuracy
±2.5 mbar (in range 700 to 1100 mbar, at 23°C±5K)
Compensation range
0 to 6500.0 mbar (programmable)
Probe dimensions
Diameter 6 mm Flow aperture approx. 10 x 3 mm
Connector
Connector colors
2 colors, light gray and dark gray, red lever
Refresh rate
0.5 seconds for all three channels
Standard baud rate
115.2 kbaud (freely selectable from 9600 baud up to 921 kbaud)
Power supply
Supply voltage
6 to 13 VDC
Current consumption
40 mA
- 31 -