7.2.2 Technical data
Operative range
Temperature (depending on sensor type) Humidity 5 to 98 % RH
Measuring quantities and ranges
Temperature
-20 to +80
℃
Accuracy 5 to +60 typical ±0.2K
℃
5 to +60 maximum ±0.4 K
℃
-20 to +80 maximum ±0.7 K
℃
Reproducibility
typical ±0.1 K
Humidity 5.0 to 98.0 % RH
Accuracy 10 to 90 % RH maximum ±2.0 % RH at 23 °C ±5 K
5 to 98 % RH maximum ±4 % RH at 23 °C ±5 K
Hysteresis
typical ±1 % RH
Atmospheric pressure
300 to 1100 mbar
Accuracy ±2.5 mbar (700 to 1100 mbar) at 23 °C ±5 K
Calculated quantities
see 7.2.1
Atmospheric pressure compensation 0 to 6500 mbar (programmable)
Refresh rate
1 second for all four channels
Connector colors
2 colors, light gray and dark gray, red lever
Standard baud rate
115.2 kbaud
Supply voltage
6 to 13 VDC
Current consumption
3 mA
Sleep mode on the device Possible (for extensions a 1-second delay is necessary)
The operating conditions are explained in Figure 7-3.
Figure 7-3 Operating conditions FHAD46C
- 16 -