6.2.1 Configurable measuring ranges
Initially the ranges for the four measuring channels can be configured from a list of
eight ranges (* factory default settings).
Description
Range
Exp. Measuring
range
Units Resolution
1. * Temperature T, t
B-01
D t
-2
-20..+80.00
+
°C
0.01 K
2. * Rel. humidity RH, U
w
B-02
D Uw
-1
5... 98.0 %H
0.1 % rH
3. * Dew point DT, t
d
B-03
D td
-1
°C
0.1 K
4. (*) Mixture MH, r mit LK
B-04
D r
-1
gk
0.1 g/kg
5. Abs. humidity AH, d
v
B-05
D dv
-1
gm
0.1 g/m
3
6. Vapor pressure VP, e
B-06
D e
-1
mb
0.1 mb
7. Enthalpy En, h mit LK
B-07
D h
-1
kJ
0.1 kJ/kg
8.(*) Atm. press. AP, p (optional)
B-08
D p
-1
300...1100.0 mb
0.1 mb
+
The measuring range depends on the sensor type.(see data sheet)
The range, the units (2 characters), and a designation are programmed automatically;
this designation comprises the familiar abbreviations listed in tables issued by the
Deutscher Wetterdienst (German Meteorological Service) and the newer symbols de-
fined in VDI/VDE 3514.
6.2.2 Technical data
Operative range
The temperature depends on the sensor type.
Humidity 5 to 98 % RH
Measuring ranges
Temperature -20 to +80 °C
Accuracy
±0.3 K at 23°C±5K
±0.4 K at 10 to 40°C
±1.3 K at -20 to 80 °C
Reproducibility: typ. ± 0.1K
Humidity 5 to 98 % RH
Accuracy ±1.8 % RH at 23 °C ±5K, 20 to 90 % RH
±2.3%RH at 23°C±5K, 10..<20%rH
Hysteresis: typ ± 1% RH.
Atmospheric pressure 300 to 1100 mbar
Accuracy ±2.5 mbar (700 to 1100 mbar) at 23°C±5K
Calculated quantities see 7.2.1
Atm. pressure compensation
0 to 6500 mbar (programmable)
Refresh rate
2 seconds for all four channels
Connector colors
2 colors, light gray and dark gray, red lever
Standard baud rate
115.2 kbaud
Supply voltage
6 to 13 VDC
Current consumption
5 mA
Sleep mode on the device
Possible (for extensions a 1s delay is necessary)
- 13 -