15.2.1 KConfigurable measuring ranges
Designation
Range
Exp
Measuring
range
Dim
Resolution
1. *flow, v 2.5 m/s
B-01
DIGI
-3
0.050 to 2.500 m/s
0.001 m/s
2. flow, v 1.0 m/s
B-02
DIGI
-3
0.050 to 1.000 m/s
0.001 m/s
3. *atmospheric pressure AP, p
B-03
DIGI
-1
300 to 1100.0 mbar
0.1 mbar
4. voltage, Volt
B-04
DIGI
-3
0.000 to 10.000 V
0.001 V
In case the flow measuring range is changed in the ALMEMO
®
plug, the correspon-
ding measuring ranges must also be changed in the sensor. For more information on
this procedure, please check the enclosed sensor documentation.
15.3 Technical data
Flow:
Measuring range:
see under 15.2.1
Resolution:
0.001 m/s
Accuracy:
± (3% of measured value + 1% of final value + 2 digits)
Nominal temperature:
23 °C +/- 2 K
Response time t63:
5 s
Atmospheric pressure:
Measuring range:
300 to 1100 mbar
Accuracy:
± 2.5 mbar (in the range of 700 to 1100 mbar) at 23°C±5K
Compensation range:
automatically in the range of 700 to 1100 mbar
Plug colors:
2 colors. light and dark grey, red levers
Refresh rate:
0.1 seconds. For both channels
Averaging time:
0.1 to 10.0 s (default value: 1.0 s)
Baud rate standard:
115.2 kBd (9600 Bd to 921 kBd selectable)
Supply voltage:
6 to 13 VDC
Current consumption:
8 mA
For further technical data, refer to the data sheet.