7.2 Configuration on a PC via the sensor menu
Figure 7-2 Sensor menu FH0D46-C
7.2.1 Configurable measuring quantities and ranges
The quantities and ranges for the four measuring channels can be configured from a
list of eight possible variants. (* factory default settings):
Designation
Quantity
Ex-
po-
nent
Measuring
range
Units Resolution
1. * Temperature T, t
B-01
D t
-2
-20..+80.00
+
°C
0.01 K
2. * Rel. humidity RH, U
w
B-02
D Uw
-1
5... 98.0 %H
0.1 % rH
3. * Dewpoint DT, t
d
B-03
D td
-1
°C
0.1 K
4. Mixture MH, r with PC
B-04
D r
-1
gk
0.1 g/kg
5. Abs. humidity AH, d
v
B-05
D dv
-1
gm
0.1 g/m
3
6. Vapor pressure VP, e
B-06
D e
-1
mb
0.1 mb
7. Enthalpy En, h mit LK
B-07
D h
-1
kJ
0.1 kJ/kg
8. * Atm. pressure AP, p
B-08
D p
-1
300...1100.0 mb
0.1 mb
+
The measuring range depends on the sensor type. (see data sheet)
The quantity, range, units (2 characters), and a comments text are programmed auto-
matically; these use the abbreviations listed in tables issued by the Deutscher Wetter-
dienst (German Meteorological Service) and the symbols more recently defined in
VDI/VDE 3514.