8. D6 temperature / humidity sensor FHAD467
Humidity sensor FHAD467 is much the same as type FHAD46 (see 7.) - with the ex-
ception that it is specially designed for use in compressed air pipes up to 16 bar. In
cases involving a pressure-dependent variable pressure compensation can be per-
formed by specifying the appropriate atmospheric pressure up to 16 bar. This amount
can also be displayed as a channel with range ´D Cp´. (see Table 8.2.1 ´with PC´)
8.1 Measuring ranges preset at our factory
Description
Range
Exp. Measuring
range
Units Resolution
1. Temperature T, t
B-01
DIGI
-2
-20..+80.00
°C
0.01 K
2. Rel. Humidity RH, U
w
B-02
DIGI
-1
5... 98.0 %H
0.1 % rH
3. Dew point DT, t
d
B-03
DIGI
-1
°C
0.1 K
The new D6 humidity ranges can be partly configured on the device itself; however, the
appropriate ALMEMO
®
standard ranges ´H DT´, ´H AH´, H VP´, ´H En´ must have
been programmed accordingly. ´DIGI´ will replace these ranges automatically with the
new ones..
Please note that new ranges ´D dv´ or ´D p´ may be lost in this process.
They can then only be restored via the PC.
8.2 Configuration on a PC via the sensor menu
- 17 -