136
U1401A User’s and Service Guide
7
Performance Tests and Calibration
8
Turn the rotary switch to
.
Connect a diode to the
U1401A input terminals in
forward bias position.
2 V
1.9 V
±1.45 mV
9
Turn the rotary switch to
mV
.
Press and hold
for
more than 1 second.
Connect K-type
thermocouple to the
U1401A input terminals.
–40 °C to
1372 °C
0 °C
±3 °C
–40 °F to
2502 °F
32 °F
±6.096 °F
Table 7-3
Input performance verification tests (continued)
Step
Test Function
Connection to 5520A
Range
5520A output
Error from
nominal 1
year
Содержание U1401A
Страница 1: ...Agilent Technologies U1401A Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401A User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401A User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401A User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401A User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401A User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401A User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401A User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401A User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401A User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401A User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401A User s and Service Guide 6 Maintenance ...