Installation
36
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
4
Place the inlet capillary tube in a beaker of de-ionized water.
Place the outlet tube of the capillary in an empty beaker.
5
From the Administration page, select SIPS from the
Configuration group. Set the following parameters:
Figure 5
The SIPS configuration page
NOTE
If you are using a SIPS 10, there is only one SIPS Arm Force field.
6
Click
Load Arm
to start the SIPS operating. Observe if the de-
ionized water in the inlet line starts to pulse or flow up the line.
If the liquid in the tube is not moving follow the steps below. If it
is moving then go to Step 9.
7
Unload the arm by clicking Unload Arm.
8
Repeat Steps 5 and 6, each time increasing the:
SIPS Arm Force/s by five units until the de-ionized water in the
line starts to pulse.
9
Increase the SIPS Arm force setting be an additional 5 units.
10
Click OK to confirm the settings.
Now condition the pump tube as described on Page 38.