Installation
28
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
WARNING
Electrical Shock – Fire Hazard
To prevent reduced safety protection or unwanted fusing, always
ensure that the marking on the fuse matches the value screen-printed
next to the fuseholder.
To check a fuse:
1
Disconnect the accessory from the mains power supply.
2
Undo the fuse cap by pressing the cap and turning it
counter-clockwise.
3
Pull the cap out carefully; the fuse should be held in the
fuseholder in the fuse cap.
4
Check that the fuse is the correct type and is not damaged. If
necessary, replace the fuse in the holder.
5
Place the fuse into the cap, push the cap in, then turn the cap
clockwise.
6
Reconnect the accessory to the mains power supply.
Always ensure both fuses are of the same type and rating before
operation.
Installing the Pump Unit and Electronic Control Module
To install the SIPS pump unit on your Agilent AA instrument, insert
the hooks on the back of the unit into the slots located on either side
of the Agilent AA instrument sample compartment.