Installation
26
Agilent Sample Introduction Pump System SIPS 10/20 User’s Guide
Installation Checklist
Use the following checklist to help you install the SIPS.
Unpack the SIPS according to the instructions and inspect the
parts for damage (refer to Page 26).
Check that the environmental conditions are correct (refer to
Page 21).
Check that the correct power supply is available and the correct
fuses are installed (refer to Page 27).
Install the pump unit and electronic control module (refer to
Page 28).
Connect the pump unit to the Agilent AA instrument or connect
the pump unit to the electronic control module (refer to Page 30).
Install the pump bands (refer to Page 32).
Prepare rinse and diluent solutions of 0.01% Triton X-100 (refer
to Page 39).
Install the SIPS diluent and pump tubing (refer to Page 33).
Calibrate the SIPS pressure arm force (refer to Page 35).
Condition the pump tubing (refer to Page 38).
Fit the dust cover (refer to Page 38).
Connect the SIPS to an autosampler (if required) (refer to Page
38).
Unpacking
Open the shipping case and remove the upper cardboard sheet. Open
each of the enclosed packages with care, inspecting the individual
parts for damage which may have occurred in transit. Any damage
should be reported to Agilent immediately. Please refer to the
accessory by part number and serial number.