Performing Time Domain Measurements
General Theory
Chapter 4
80
General Theory
The Agilent multiport test system combined with Agilent’s vector network analyzers and their
companion multiport test sets perform measurements in the frequency domain by sweeping
an RF signal and measuring the RF responses of a device under test (DUT). The software can
also mathematically transform these frequency domain data S parameters into their time
domain counterparts and display them in either their step, impedance or their
impulse/response modes.
In a linear network, the Fourier Transform describes the relationship between a frequency
domain measurement and its corresponding time domain response in detail. Therefore, given
the measured frequency domain response of a DUT, it is possible to determine its time domain
response mathematically by performing an inverse Fourier Transform. And the software
accomplishes it’s frequency domain transformation to time domain by utilizing this inverse
chirp Z Fourier transform
1
.
As shown in the following illustration, the type of information that can be observed in time
domain mode is quite different than the information that can be observed in frequency
domain mode. If the network is thought of in terms of its equivalent circuit model, then the
frequency domain response describes the composite behavior of all of the circuit elements at
any given operating frequency.
By contrast, the time-domain response shows the contribution of each individual circuit
element. Since there is a direct relationship between time and distance, this mode allows each
element to be separated spatially. With an understanding of the unique signature
characteristics of different circuit elements, this view of the DUT can provide considerable
insight into the device.
1. The advantage of the chirp z-transform is that it enables calculation of the sample of the
z-transform equally spaced over an arc or a spiral contour with an arbitrary starting point
and arbitrary frequency range. In contrast, the frequency range of the discrete Fourier
Transform is strictly related to the sampling frequency.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...