Using the Multiport Test System
How to Set Up the Multiport Test System Hardware
Chapter 3
38
To Set the GPIB Addresses
The test set uses GPIB to communicate with system components. The GPIB address switch is
on the back panel of the test set; it is set to “18” at the factory. While this will work in most
situations, the address can be changed if required to avoid address conflicts with other
equipment on the same bus. If you change the address on the test set, make sure you also
change it in the software. (See
“Setup > Hardware…” on page 124
GPIB addresses are five-bit binary numbers set with the switches labeled “16” through “1”,
with “1” being the least significant digit.
“T”, “L”, and “S” correspond to Talk, Listen, and Status. Set them as shown (off).
The example below shows the default setting, “18”, set by turning on switches “16” and “2”.
NOTE
After changing a GPIB address setting, on any of the system equipment, turn
the test set AC power off and back on establish the new GPIB address.
Pull down
Setup > Hardware…
to set the test set GPIB address in the software. See
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...