Menu Reference
Calibrate Menu
Chapter 5
134
Click
Show Sample of Cal Freqs
to display a sample of calibration frequencies from the
current full-sweep calibration.
The
CURRENT CAL SETTINGS
line shows the Start and Stop frequencies and the number
of points in the current full-sweep calibration.
In the
START:
and
STOP:
fields, enter the start and stop frequencies for the new calibration
subset. Changing these fields allow you to sweep a narrower span while the calibration subset
option is active.
In the
SKIP PTS:
field, enter the number of points to skip in the full calibration to create the
new calibration subset.
The
SUBSET CAL SETTINGS
line shows the Start and Stop frequencies and the number of
points in the new calibration subset.
Click
Create Cal Subset
to create the new subset of the current full-sweep calibration set.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...