Menu Reference
Setup Menu
Chapter 5
120
Setup > Test Port Impedance…
Pull down
Setup > Test Port Impedance…
to open the
Port Impedance
dialog box and set
reference impedances.
By default all four ports are set to 50
Ω
impedance. Enter new reference impedances in the
port boxes as desired.
This feature does not affect the impedance presented to the DUT by the system hardware.
Rather, it mathematically transforms the measured data to show how the performance of a
linear device would change in a non-50
Ω
measurement system. Impedance transforms can be
specified either before or after measurement. Only the current data set is affected.
The reference impedance can be specified as resistive, or as an RL or RC circuit. The
impedance model is used to specify whether the selected impedance is the effective
source/load impedance, or whether it is an estimate of the DUT port impedance. One is the
complex conjugate of the other.
Scroll through the
Recent Configs
box to select a configuration from one of the last ten
reference impedance configurations.
Click
Set All Ports to 50 ohms
to return to the default configuration.
Click
OK
to accept the configuration and exit the dialog box.
Click
Cancel
to ignore any changes and exit the dialog box.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...