Using the Multiport Test System
How to Use Frequency Zoom
Chapter 3
55
The
STOP
: line shows the ending frequency and a two-line confirmation menu displays:
7. Click on
Zoom
to confirm selected settings. The part of the trace between the start and
stop points is zoomed.
8. To restore the original measurement range, right click in the gray area outside of the plot
to display the display properties menu and select
Restore Full Sweep Data
from the
menu.
If frequency zooming has been applied to the data, the
+Data
button appears in the indicator
bar:
If a measurement is made after the frequency axis is zoomed, only the points within the zoom
range are measured. This creates a new set of calibration data adjusted to the new frequency
range.
As soon as the measurement starts, the
+Cal
button appears in the indicator bar to show that
the calibration in use is a subset of the full calibration available:
At the end of the measurement, the
+Data
button disappears because all available
measurement points have been displaced.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...