Instrument Specifications
* The listed conditions apply to instrument operation. Conditions required for assay performance may
differ.
Table 1
Technical specifications of the instrument
System Component
Specification
Thermal cycler module
Minimum temperature of thermal block
4°C
Maximum temperature of thermal block
99°C
Heater/Shaker/Magnet module
Heater
Temperature maximum
75°C
Shaker
Maximum speed (rpm)
1800 rpm ±5%
Chiller module
Temperature range
4–12°C
Power inlet
AC voltage
100–240 VAC
AC frequency
50/60 Hz
Maximum power
1000 W
System
Acoustic sound pressure
70 dBA
Dimensions, instrument door closed (Height × Depth × Width)
71 cm × 72 cm × 62 cm
Dimensions, instrument door open (Height × Depth × Width)
107 cm × 72 cm × 62 cm
Weight
95 kg
Environmental conditions*
Temperature
Operating: 15°C to 25°C
Transport and storage: –40°C to 70°C
Humidity
Operating: 30% to 70%, non-condensing
Altitude
2000 m