Magnis and Magnis Dx NGS Prep System User Guide
70
Figure 47
Hardware Usage Tracking screen
UV Tube
Next to this check box is displayed the number of hours, minutes, and seconds of usage for the
UV tube. Marking the check box enables the Reset button
Reset
This button resets the usage tracker for the UV tube to zero. Once reset, the usage is displayed as
0h 00m 00s
.
This action should only be performed by an Agilent engineer or Agilent authorized
service provider following replacement of the UV tube.
Close
This button returns you to the Settings screen.
Instrument Diagnostic screen
Purpose: To run diagnostic tests on the instrument and access reports for previous diagnostic
tests and instrument health checks. See instructions in
“Perform instrument diagnostic tests”
To open: From the Home screen, press
Settings
. On the Settings screen press
Self Diagnostic
.
NOTE
The lifespan of the UV tube is 630 hours. After 630 hours of usage, when a
decontamination cycle is initiated, the system notifies you to replace the UV
tube.