Magnis and Magnis Dx NGS Prep System User Guide
41
Decontaminating with UV Light
Run a decontamination “quick cycle”
The Magnis/Magnis Dx NGS Prep System includes a UV tube that can be used to decontaminate
the surfaces of the instrument deck. The
quick cycle
is a 30-minute decontamination. Agilent
recommends running the quick cycle before each protocol run.
1
Verify that the instrument deck is cleared of all labware, then make sure that the instrument
door is closed.
2
From the Home screen, press
Decontamination
.
The
Figure 24
Decontamination screen with Quick cycle selected
3
At the top of the screen, select
Quick cycle
, as shown in
4
Press
Start
.
The decontamination cycle begins and the screen displays a countdown of the time remaining.
At the completion of the cycle, the UV light turns off and the instrument remains idle.
If needed, press
Abort
at any time to stop the decontamination cycle.
WARNING
Do not look directly at the UV light while decontamination is in progress.