Magnis and Magnis Dx NGS Prep System User Guide
77
Figure 55
Other Settings screen
Verify teach points during Instrument Health Check
When this check box is marked, each time the system is powered on, the first IHC that it performs
includes verification of the teach point positions. If the system identifies a mis-alignment with the
teach points, it displays an error message instructing you to run auto-teaching. See
Auto-Teaching and Teach-Point Verification”
Close