Decontamination screen
Purpose: To perform UV decontamination of the instrument deck. See instructions in
“Decontaminating with UV Light”
To open: From the Home screen, press
Decontamination
.
Figure 60
Decontamination screen
Quick cycle / Extended cycle
Select between the quick cycle and extended cycle decontamination procedures.
The quick cycle takes 30 minutes to complete. Agilent recommends running the quick cycle prior
to each protocol run.
The extended cycle takes 2 hours to complete. Agilent recommends running the extended cycle in
the event of a spill or leak that resulted in potential contamination of the instrument deck. At the
end of the extended cycle, the instrument automatically powers off. The system only permits
running the extended cycle once every 7 days to avoid excessive UV light exposure on the deck.
Time Remaining
The time displayed is the time remaining (hh:mm:ss) in the selected decontamination cycle.
Start
This button initiates the decontamination cycle.
Close
This button is available before the decontamination cycle is started. It returns you to the Home
screen.
Abort
This button is available while a decontamination cycle is in-progress. It stops the cycle and turns
off the UV light.