Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:TRIGger:DURation:GREaterthan
Command Syntax
:TRIGger:DURation:GREaterthan <greater_than_time>[<suffix>]
<greater_than_time> ::= minimum trigger duration in seconds
in NR3 format
<suffix> ::= {s | ms | us | ns | ps }
The :TRIGger:DURation:GREaterthan command sets the minimum duration
for the defined pattern when :TRIGger:DURation:QUALifier is set to
GREaterthan. The command also sets the timeout value when the
:TRIGger:DURation:QUALifier is set to TIMeout.
Query Syntax
:TRIGger:DURation:GREaterthan?
The :TRIGger:DURation:GREaterthan? query returns the minimum duration
time for the defined pattern.
Return Format
<greater_than_time><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
Содержание InfiniiVision 7000A Series
Страница 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Страница 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 652: ...652 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 5 Commands by Subsystem...
Страница 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Страница 784: ...784 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 10 Synchronizing Acquisitions...
Страница 810: ...810 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 11 More About Oscilloscope Commands...
Страница 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...