Obsolete and Discontinued Commands
7
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
699
:DISPlay:CONNect
Command Syntax
:DISPlay:CONNect <connect>
<connect> ::= {{ 1 | ON} | {0 | OFF}}
The :DISPlay:CONNect command turns vectors on and off. When vectors
are turned on, the oscilloscope displays lines connecting sampled data
points. When vectors are turned off, only the sampled data is displayed.
Query Syntax
:DISPlay:CONNect?
The :DISPlay:CONNect? query returns the current state of the vectors
setting.
Return Format
<connect><NL>
<connect> ::= {1 | 0}
See Also
•
N O T E
The :DISPlay:CONNEct command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :DISPlay:VECTors command (see
) instead.
Содержание InfiniiVision 7000A Series
Страница 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Страница 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 652: ...652 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 5 Commands by Subsystem...
Страница 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Страница 784: ...784 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 10 Synchronizing Acquisitions...
Страница 810: ...810 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 11 More About Oscilloscope Commands...
Страница 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...