492
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:TRIGger:DURation:PATTern
Command Syntax
:TRIGger:DURation:PATTern <value>, <mask>
<value> ::= integer or <string>
<mask> ::= integer or <string>
<string> ::= "0xnnnnnn"; n ::= {0,..,9 | A,..,F}
The :TRIGger:DURation:PATTern command defines the specified duration
pattern resource according to the value and the mask. For both <value>
and <mask>, each bit corresponds to a possible trigger channel. The bit
assignments vary by instrument:
Set a <value> bit to "0" to set the pattern for the corresponding channel to
low. Set a <value> bit to "1" to set the pattern to high.
Set a <mask> bit to "0" to ignore the data for the corresponding channel.
Only channels with a "1" set on the appropriate mask bit are used.
Query Syntax
:TRIGger:DURation:PATTern?
The :TRIGger:DURation:PATTern? query returns the pattern value.
Return Format
<value>, <mask><NL>
<value> ::= a 32-bit integer in NR1 format.
<mask> ::= a 32-bit integer in NR1 format.
See Also
•
"Introduction to :TRIGger Commands"
•
Oscilloscope Models
Value and Mask Bit Assignments
4 16 digital channels (mixed-signal)
Bits 0 through 15 - digital channels 0 through
15. Bits 16 through 19 - analog channels 1
through 4.
2 16 digital channels (mixed-signal)
Bits 0 through 15 - digital channels 0 through
15. Bits 16 and 17 - analog channels 1 and 2.
4 analog channels only
Bits 0 through 3 - analog channels 1 through 4.
Bit 4 - external trigger.
2 analog channels only
Bits 0 and 1 - analog channels 1 and 2. Bit 4 -
external trigger.
Содержание InfiniiVision 7000A Series
Страница 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Страница 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 652: ...652 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 5 Commands by Subsystem...
Страница 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Страница 784: ...784 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 10 Synchronizing Acquisitions...
Страница 810: ...810 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 11 More About Oscilloscope Commands...
Страница 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...