:MTESt:RMODe:FACTion:MEASure
Command Syntax
:MTESt:RMODe:FACTion:MEASure <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:MEASure command sets measuring only mask
failures on or off.
When ON, measurements and measurement statistics run only on
waveforms that contain a mask violation; passing waveforms do not affect
measurements and measurement statistics.
This mode is not available when the acquisition mode is set to Averaging.
Query Syntax
:MTESt:RMODe:FACTion:MEASure?
The :MTESt:RMODe:FACTion:MEASure? query returns the current mask
failure measure setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
•
"Introduction to :MTESt Commands"
•
•
•
Содержание InfiniiVision 7000A Series
Страница 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Страница 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Страница 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Страница 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Страница 652: ...652 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 5 Commands by Subsystem...
Страница 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Страница 784: ...784 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 10 Synchronizing Acquisitions...
Страница 810: ...810 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 11 More About Oscilloscope Commands...
Страница 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...