54
6000 Series Oscilloscope User’s Guide
1
Getting Started
To compensate the oscilloscope probes
You should compensate your oscilloscope probes to match their
characteristics to the oscilloscope’s channels. A poorly
compensated probe can introduce measurement errors.
1
Perform the procedure
“To verify basic oscilloscope
operation”
on page 52.
2
Use a nonmetallic tool to adjust the trimmer capacitor on the
probe for the flattest pulse possible. The trimmer capacitor is
located on the probe BNC connector.
3
Connect probes to all other oscilloscope channels (channel 2
of a 2-channel oscilloscope, or channels 2, 3, and 4 of a
4-channel oscilloscope). Repeat the procedure for each
channel. This matches each probe to each channel.
The process of compensating the probes serves as a basic test to
verify that the oscilloscope is functional.
comp.cdr
Perfectly compensated
Over compensated
Under compensated