Sharing one probe ground with many probes forces all the
current that flows into each probe to return through the same
common ground inductance of the probe whose ground return
is used. The result is increased current (
di
) in the above
equation, and, depending on the transition time (
dt
), the
common mode voltage may increase to a level that causes false
data generation.
Figure 52
Common Mode Input Voltage Model
In addition to the common mode voltage,
longer ground returns
also degrade the pulse fidelity of the probe system. Rise time is
increased, and ringing, due to the undamped LC circuit at the
input of the probe, is also increased. Because the digital
channels display reconstructed waveforms, they do not show
ringing and perturbations. You will not find ground problems
through examination of the waveform display. In fact, it is likely
you will discover the problem through random glitches or
inconsistent data measurements. Use the analog channels to
view ringing and perturbations.
Z
in
Z
in
i
n
i
2
+i
2
i
1
i
1
+i
n
+i
n
Z
in
Probe 1
L (GND)
Probe 2
Probe N
V
n
(Common Mode)
Probe
Ground
Содержание InfiniiVision 6000 Series
Страница 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Страница 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Страница 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Страница 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Страница 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Страница 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Страница 362: ...362 6000 Series Oscilloscope User s Guide 9 ...