166
6000 Series Oscilloscope User’s Guide
4
Triggering the Oscilloscope
Data needs to be stable during the whole high clock cycle or
it will be interpreted as a start or stop condition (data
transitioning while the clock is high).
6
Return to the previous menu; then, press the
Trigger:
softkey
to select one of the following I
2
C trigger conditions:
•
Start Condition
— The oscilloscope triggers when SDA data
transitions from high to low while the SCL clock is high.
For triggering purposes (including frame triggers), a
restart is treated as a start condition.
•
Stop Condition
— The oscilloscope triggers when data (SDA)
transitions from low to high while the clock (SCL) is high.
•
Missing Acknowledge
— The oscilloscope triggers when SDA
data is high during any Ack SCL clock bit.
•
Address with no Ack
— The oscilloscope triggers when the
acknowledge of the selected address field is false. The R/W
bit is ignored.
•
Restart
— The oscilloscope triggers when another start
condition occurs before a stop condition.
SDA
SCL
Address
R/
Ack
Data
Ack
Stop
Condition
Start
Condition
Содержание InfiniiVision 6000 Series
Страница 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Страница 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Страница 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Страница 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Страница 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Страница 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Страница 362: ...362 6000 Series Oscilloscope User s Guide 9 ...