Power and Environmental Conditions
9
6000 Series Oscilloscope User’s Guide
359
Environmental Conditions
Environment
Indoor use only.
Ambient temperature
Operating –10 °C to +55 °C;
non-operating –51 °C to +71 °C
Humidity
Operating 95% RH at 40 °C for 24 hr; non-operating
90% RH at 65 °C for 24 hr
Altitude
Operating to 4,570 m (15,000 ft); non-operating to
15,244 m (50,000 ft)
Overvoltage Category
This product is intended to be powered
by MAINS that comply to Overvoltage Category II, which is
typical of cord-and-plug connected equipment.
Pollution Degree
The 6000 Series Oscilloscope may be operated
in environments of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree Definitions
Pollution Degree 1: No pollution or
only dry, non-conductive pollution occurs. The pollution has no
influence. Example: A clean room or climate controlled office
environment.
Pollution Degree 2. Normally only dry non-conductive pollution
occurs. Occasionally a temporary conductivity caused by
condensation may occur. Example: General indoor
environment.
Pollution Degree 3: Conductive pollution occurs, or dry,
non-conductive pollution occurs which becomes conductive due
to condensation which is expected. Example: Sheltered outdoor
environment.
Содержание InfiniiVision 6000 Series
Страница 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Страница 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Страница 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Страница 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Страница 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Страница 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Страница 362: ...362 6000 Series Oscilloscope User s Guide 9 ...