Making Measurements
5
6000 Series Oscilloscope User’s Guide
229
resolve two closely space frequencies. A good way to test the
ability of the FFT to resolve two closely spaced frequencies is to
examine the sidebands of an amplitude modulated sine wave.
For the best vertical accuracy on peak measurements:
•
Make sure the probe attenuation is set correctly. The probe
attenuation is set from the Channel menu if the operand is a
channel.
•
Set the source sensitivity so that the input signal is near full
screen, but not clipped.
•
Use the Flat Top window.
•
Set the FFT sensitivity to a sensitive range, such as
2 dB/division.
For best frequency accuracy on peaks:
•
Use the Hanning window.
•
Use Cursors to place an X cursor on the frequency of
interest.
•
Adjust frequency span for better cursor placement.
•
Return to the Cursors menu to fine tune the X cursor.
For more information on the use of FFTs please refer to Agilent
Application Note 243, ”The Fundamentals of Signal Analysis” at
http://cp.literature.agilent.com/litweb/pdf/5952-8898E.pdf
.
Additional information can be obtained from Chapter 4 of the
book “Spectrum and Network Measurements” by Robert A.
Witte.
Содержание InfiniiVision 6000 Series
Страница 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Страница 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Страница 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Страница 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Страница 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Страница 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Страница 362: ...362 6000 Series Oscilloscope User s Guide 9 ...