Displaying Data
6
6000 Series Oscilloscope User’s Guide
269
Figure 46
128 Averages used to reduce random noise
Realtime Sampling Option
Realtime sampling specifies that the oscilloscope produce the
waveform display from samples collected during one trigger
event (that is, one acquisition).
Use Realtime sampling to capture infrequent triggers, unstable
triggers, or complex changing waveforms, such as eye diagrams.
Realtime sampling can be turned on in
Normal
,
Peak Detect
, or
High Resolution
acquisition modes. It cannot be turned on when
the acquisition mode is
Averaging
.
When Realtime sampling is on (as in the default setup):
•
When less than 1000 samples can be collected in the time
spanned by the screen, a sophisticated reconstruction filter
is used to fill in and enhance the waveform display.
•
If you press the
Stop
key, and pan and zoom through the
waveform using the Horizontal and Vertical controls, only the
last trigger’s acquisition will be displayed.
When Realtime sampling is off:
Содержание InfiniiVision 6000 Series
Страница 1: ...Agilent Technologies Agilent InfiniiVision 6000 Series Oscilloscopes User s Guide ...
Страница 60: ...60 6000 Series Oscilloscope User s Guide 1 Getting Started ...
Страница 126: ...126 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals ...
Страница 292: ...292 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data ...
Страница 314: ...314 6000 Series Oscilloscope User s Guide 6 Displaying Data ...
Страница 354: ...354 6000 Series Oscilloscope User s Guide 8 Reference ...
Страница 362: ...362 6000 Series Oscilloscope User s Guide 9 ...