18
This section documents performance test procedures. Performance verification for the
products covered by this manual consists of three main steps:
• Performing the internal product self-tests to ensure that the measurement system is
functioning properly
• Calibrating the product
• Testing the product to ensure that it is performing to specification
Performance Test Interval
The procedures in this section may be performed for incoming inspection and should
be performed periodically to verify that the oscilloscope is operating within
specification. The recommended test interval is once per year or after 2000 hours of
operation. Performance should also be tested after repairs or major upgrades.
Performance Test Record
A test record form is provided at the end of this section. This record lists performance
tests, test limits and provides space to record test results.
Test Order
The tests in this section may be performed in any order desired. However, it is
recommended to conduct the tests in the order presented in this manual as this
represents an incremental approach to performance verification. This may be useful if
you are attempting to troubleshoot a suspected problem.
Test Equipment
Lists of equipment needed to conduct each test are provided for each test procedure.
The procedures are written to minimize the number and types of oscilloscopes and
accessories required. The oscilloscopes in these lists are ones that are currently available
for sale by Agilent at the time of writing this document. In some cases, the test
procedures use features specific to the oscilloscopes in the recommended equipment
list. However, with some modification to the test procedures, oscilloscopes, cables and
accessories that satisfy the critical specifications in these lists may be substituted for
the recommended models with some modification to the test procedures.
Contact Agilent Technologies for more information about the Agilent products in these
lists.
Содержание Infiniium 90000 A
Страница 4: ...4...
Страница 13: ...2 To run the self calibration 15 Calibration...
Страница 14: ...14 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 50: ...Chapter 3 Testing Performance Performance Test Record 50...
Страница 58: ...Chapter 4 Troubleshooting Power Supply Trouble Isolation 58 Figure 4 4...
Страница 71: ...Chapter 4 Troubleshooting Display Trouble Isolation 71 Display Trouble Isolation...
Страница 88: ...Chapter 4 Troubleshooting To check probe power outputs 88...
Страница 111: ...Chapter 5 Replacing Assemblies To remove and replace the acquisition boards backplane assembly 111 Figure 5 36...
Страница 126: ...Chapter 5 Replacing Assemblies To remove and replace the USB or GPIB port 126...
Страница 129: ...Chapter 6 Replaceable Parts Exploded Views 129 Exploded Views Front Frame and Front Panel...
Страница 130: ...Chapter 6 Replaceable Parts Exploded Views 130 Fan and Acquisition Assembly...
Страница 131: ...Chapter 6 Replaceable Parts Exploded Views 131 Power Supply and PC Motherboard...
Страница 132: ...Chapter 6 Replaceable Parts Exploded Views 132 Sleeve and Accessory Pouch...
Страница 136: ...Chapter 6 Replaceable Parts Replaceable Parts List 136...
Страница 138: ...Chapter 7 Theory of Operation 138 Figure 7 1 Oscilloscope Block Diagram...
Страница 140: ...Chapter 7 Theory of Operation Block Level Theory 140 Figure 7 2 AcquisitionBoards BackplaneBoard...
Страница 146: ...146 Figure 7 6 Acquisition board block diagram...