Chapter 5: Replacing Assemblies
To return the oscilloscope to Agilent Technologies for service
91
W A R N I N G
SHOCK HAZARD!
When the bulk 12V power supply is removed from the oscilloscope, two AC leads are exposed.
Therefore, the power cable should be disconnected from the bulk 12V power supply before the
supply is removed. Caution should also be exercised to not contact these leads as severe shock
could result.
To return the oscilloscope to Agilent Technologies for service
Before shipping the oscilloscope to Agilent Technologies, contact your nearest Agilent
Technologies oscilloscope Support Center (or Agilent Technologies Service Center if outside the
United States) for additional details.
1
Write the following information on a tag and attach it to the oscilloscope.
• Name and address of owner
• Oscilloscope model numbers
• Oscilloscope serial numbers
• Description of the service required or failure indications
2
Remove all accessories from the oscilloscope.
Accessories include all cables. Do not include accessories unless they are associated with the
failure symptoms.
3
Protect the oscilloscope by wrapping it in plastic or heavy paper.
4
Pack the oscilloscope in foam or other shock absorbing material and place it in a strong
shipping container.
You can use the original shipping materials or order materials from an Agilent Technologies Sales
Office. If neither are available, place 8 to 10 cm (3 to 4 inches) of shock-absorbing material around
the oscilloscope and place it in a box that does not allow movement during shipping.
5
Seal the shipping container securely.
6
Mark the shipping container as FRAGILE.
In any correspondence, refer to oscilloscope by model number and full serial number.
Содержание Infiniium 90000 A
Страница 4: ...4...
Страница 13: ...2 To run the self calibration 15 Calibration...
Страница 14: ...14 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 50: ...Chapter 3 Testing Performance Performance Test Record 50...
Страница 58: ...Chapter 4 Troubleshooting Power Supply Trouble Isolation 58 Figure 4 4...
Страница 71: ...Chapter 4 Troubleshooting Display Trouble Isolation 71 Display Trouble Isolation...
Страница 88: ...Chapter 4 Troubleshooting To check probe power outputs 88...
Страница 111: ...Chapter 5 Replacing Assemblies To remove and replace the acquisition boards backplane assembly 111 Figure 5 36...
Страница 126: ...Chapter 5 Replacing Assemblies To remove and replace the USB or GPIB port 126...
Страница 129: ...Chapter 6 Replaceable Parts Exploded Views 129 Exploded Views Front Frame and Front Panel...
Страница 130: ...Chapter 6 Replaceable Parts Exploded Views 130 Fan and Acquisition Assembly...
Страница 131: ...Chapter 6 Replaceable Parts Exploded Views 131 Power Supply and PC Motherboard...
Страница 132: ...Chapter 6 Replaceable Parts Exploded Views 132 Sleeve and Accessory Pouch...
Страница 136: ...Chapter 6 Replaceable Parts Replaceable Parts List 136...
Страница 138: ...Chapter 7 Theory of Operation 138 Figure 7 1 Oscilloscope Block Diagram...
Страница 140: ...Chapter 7 Theory of Operation Block Level Theory 140 Figure 7 2 AcquisitionBoards BackplaneBoard...
Страница 146: ...146 Figure 7 6 Acquisition board block diagram...