![Agilent Technologies Infiniium 90000 A Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/agilent-technologies/infiniium-90000-a/infiniium-90000-a_service-manual_2868672021.webp)
Chapter 3: Testing Performance
Offset Accuracy Test
21
Offset Accuracy Test
C A U T I O N
Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
Specifications
Equipment Required
The offset accuracy specification has two terms ±(offset gain + zero error). The offset gain
specification is ±2% of channel offset and the zero error specification is ±1% of full scale. The
offset accuracy test procedure tests each of these terms individually.
Procedure
Zero Error Test
1
Disconnect all cables from the scope channel inputs.
2
Press Default Setup, then configure the scope as follows:
a
Pull down the Setup menu and select Acquisition.
Let the oscilloscope warm up before testing
The oscilloscope under test must be warmed up (with the oscilloscope application running) for at least
30 minutes prior to the start of any performance test.
Offset Accuracy
≤
3.5 V: ±(2% of channel 1% of full scale + 1 mV)
Full scale is defined as 8 vertical divisions. Magnification is used below 5 mV/div. Below 5
mV full scale is defined as 40 mV. The major scale settings are 5 mV, 10 mV, 20 mV,
50 mV, 100 mV, 200 mV, 500 mV and 1 V.
Description
Critical Specifications
Recommended
Model/
Part Numbers
Digital Multimeter
DC voltage measurement accuracy better than ±0.1% of reading
Agilent 34401A or
Agilent 3458A
Cable Assembly
(2 required)
50
Ω
characteristic impedance, BNC (m) connectors
Agilent 8120-1840
Adapter
BNC Tee (m)(f)(f)
Agilent 1250-0781
Adapter
BNC (f) to dual banana
Agilent 1251-2277
Содержание Infiniium 90000 A
Страница 4: ...4...
Страница 13: ...2 To run the self calibration 15 Calibration...
Страница 14: ...14 Calibration This chapter provides self calibration procedures for the oscilloscope...
Страница 50: ...Chapter 3 Testing Performance Performance Test Record 50...
Страница 58: ...Chapter 4 Troubleshooting Power Supply Trouble Isolation 58 Figure 4 4...
Страница 71: ...Chapter 4 Troubleshooting Display Trouble Isolation 71 Display Trouble Isolation...
Страница 88: ...Chapter 4 Troubleshooting To check probe power outputs 88...
Страница 111: ...Chapter 5 Replacing Assemblies To remove and replace the acquisition boards backplane assembly 111 Figure 5 36...
Страница 126: ...Chapter 5 Replacing Assemblies To remove and replace the USB or GPIB port 126...
Страница 129: ...Chapter 6 Replaceable Parts Exploded Views 129 Exploded Views Front Frame and Front Panel...
Страница 130: ...Chapter 6 Replaceable Parts Exploded Views 130 Fan and Acquisition Assembly...
Страница 131: ...Chapter 6 Replaceable Parts Exploded Views 131 Power Supply and PC Motherboard...
Страница 132: ...Chapter 6 Replaceable Parts Exploded Views 132 Sleeve and Accessory Pouch...
Страница 136: ...Chapter 6 Replaceable Parts Replaceable Parts List 136...
Страница 138: ...Chapter 7 Theory of Operation 138 Figure 7 1 Oscilloscope Block Diagram...
Страница 140: ...Chapter 7 Theory of Operation Block Level Theory 140 Figure 7 2 AcquisitionBoards BackplaneBoard...
Страница 146: ...146 Figure 7 6 Acquisition board block diagram...