FR2 AED Specifications
ECG Analysis Performance
Shockable Rhythms
Ventricular fibrillation (VF) and certain ventricular tachycardias, including
ventricular flutter and polymorphic ventricular tachycardia (VT). The
H
EARTSTREAM
FR2 uses multiple parameters to determine if a rhythm is
shockable.
NOTE: For safety reasons, some very low-amplitude or low-frequency rhythms
may not be interpreted as shockable VF rhythms. Also, some VT rhythms may
not be interpreted as shockable rhythms.
Asystole
On detection of asystole, provides CPR prompt at programmed interval.
Pacemaker Detection
On detection of a pacemaker (in advanced mode only), provides screen display of
PACEMAKER DETECTED alert, and M3860A includes pacemaker artifact in ECG
display. In both models, pacemaker artifact is removed from the signal for rhythm
analysis.
CATEGORY
NOMINAL SPECIFICATIONS
RHYTHM CLASS
ECG TEST
SAMPLE
a
SIZE
NOMINAL SPECIFICATIONS
Shockable Rhythm —
ventricular fibrillation
300
Meets AAMI DF39 requirement and AHA recommendation
b
(sensitivity >90%) for adult defibrillation
Shockable Rhythm —
ventricular tachycardia
100
Meets AAMI DF39 requirement and AHA recommendation
b
(sensitivity >75%) for adult defibrillation
Non-Shockable Rhythm —
Normal Sinus Rhythm
300
Meets AAMI DF39 requirement (specificity > 95%) and AHA
recommendation
b
(specificity >99%) for adult defibrillation
Non-Shockable Rhythm —
Asystole
100
Meets AAMI DF39 requirement and AHA recommendation
b
(specificity >95%) for adult defibrillation
Non-Shockable Rhythm —
All other non-shockable
rhythms
450
Meets AAMI DF39 requirement and AHA recommendation
b
(specificity >95%) for adult defibrillation
a. From Agilent Technologies Heartstream Operation ECG rhythm databases.
b. American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy. Auto-
matic External Defibrillators for Public Access Use: Recommendations for Specifying and Reporting
Arrhythmia Analysis Algorithm Performance, Incorporation of New Waveforms, and Enhancing Safety.
American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy.
Circu-
lation
1997;95:1677-1682.
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...