Safety Considerations
5-2
Clinical and Safety Considerations
Safety Considerations
You should be aware of the safety concerns listed here when you use the
H
EARTSTREAM
FR2. Read them carefully. You will also see some of these messages in
other parts of this User’s Guide. The messages are labeled Danger, Warning, or
Caution.
DANGER — immediate hazards that will result in personal injury or death.
WARNING — conditions, hazards, or unsafe practices that can result in
serious personal injury or death.
CAUTION — conditions, hazards, or unsafe practices that can result in
minor personal injury, damage to the H
EARTSTREAM
FR2, or loss of data
stored in the device.
These safety considerations are divided into four groups: safety concerns about the
H
EARTSTREAM
FR2 in general use, defibrillation, monitoring, and maintenance
activities.
The dangers, warnings, and cautions listed in the following tables apply to both the
model M3860A and the model M3861A H
EARTSTREAM
FR2, unless otherwise noted.
General Dangers, Warnings, and Cautions
SAFETY LEVEL
POSSIBLE SHOCK OR FIRE HAZARD, OR EXPLOSION
DANGER
THERE IS A POSSIBILITY OF EXPLOSION IF THE H
EARTSTREAM
FR2 IS USED IN THE PRESENCE
OF FLAMMABLE ANESTHETICS OR CONCENTRATED OXYGEN.
DANGER
THE H
EARTSTREAM
FR2 HAS NOT BEEN EVALUATED OR APPROVED FOR USE IN HAZARDOUS
LOCATIONS AS DEFINED IN THE NATIONAL ELECTRICAL CODE (ARTICLES 500-503). IN
ACCORDANCE WITH THE IEC CLASSIFICATIONS (SECTION 5.5.), THE H
EARTSTREAM
FR2 IS
NOT TO BE USED IN THE PRESENCE OF FLAMMABLE SUBSTANCE/AIR MIXTURES.
DANGER
DO NOT RECHARGE THE M3863A BATTERY.
WARNING
Use the H
EARTSTREAM
FR2 only as described in this User’s Guide. Improper use of the
H
EARTSTREAM
FR2 can cause death or injury. Do not press the Shock button if the defibrillation
pads are touching each other or are open and exposed.
CAUTION
Hazardous electrical output. The H
EARTSTREAM
FR2 is for use only by qualified personnel.
CAUTION
Do not immerse any portion of the H
EARTSTREAM
FR2 in water or other fluids. Do not allow fluids to enter
the H
EARTSTREAM
FR2. Avoid spilling any fluids on the H
EARTSTREAM
FR2 or accessories. Spilling fluids into
the H
EARTSTREAM
FR2 may damage it or present a fire or shock hazard. Do not sterilize the H
EARTSTREAM
FR2 or accessories.
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...