M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
B-1
B
Technical Specifications
The specifications for the H
EARTSTREAM
FR2 provided in this chapter apply to both
the M3860A and M3861A, unless otherwise noted.
H
EARTSTREAM
FR2 AED Specifications
Physical
Environmental
CATEGORY
NOMINAL SPECIFICATIONS
Size
2.6" high x 8.6" wide x 8.6" deep (6.6 cm x 21.8 cm x 21.8 cm).
Weight
Approximately 4.7 lbs (2 kg) with battery installed.
CATEGORY
NOMINAL SPECIFICATIONS
Operating Temperature
and Humidity
32° to 122° F (0° to 50° C).
0% to 95% relative humidity (non-condensing).
Standby Temperature
and Humidity
32° to 109° F (0° to 43° C).
0% to 75% relative humidity (non-condensing).
Applies to H
EARTSTREAM
FR2 with battery installed and stored with defibrillation
pads.
Altitude
Meets MIL-810E 500.3, Procedure II (-500 feet to 15,000 feet).
Shock/Drop Abuse
Tolerance
Meets MIL-STD-810E 516.4, Procedure IV (after a 1 meter drop to any edge,
corner, or surface, in standby mode).
Vibration
Meets MIL-STD-810E 514.4-17.
Sealing
With data card tray and battery installed, meets IEC 529 class IP54.
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...