Troubleshooting Guide
M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
4-17
4
Voice and screen prompts:
NO SHOCK DELIVERED
The patient impedance is not
appropriate for the FR2 to deliver a
biphasic shock.
• Make sure the defibrillation pads are
correctly positioned on the patient.
• Make sure the defibrillation pads
connector is completely inserted in
the connector socket.
• Press the defibrillation pads firmly to
the patient's chest.
• Replace the defibrillation pads if
necessary.
Voice prompt:
SHOCK BUTTON
NOT PRESSED
A shock has been advised but not
delivered within 30 seconds. (The
FR2 has been disarmed.)
When next prompted, press the Shock
button to deliver a shock.
OTHER SIGNAL
POSSIBLE CAUSE
RECOMMENDED ACTION
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...