Testing
M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
4-5
4
Testing
The H
EARTSTREAM
FR2 has several ways of testing itself and alerting you if it finds a
problem. In addition to the selftest performed each time a battery is installed, the
H
EARTSTREAM
FR2 also automatically performs periodic selftests daily.
NOTE:
The FR2 selftests are designed to check that the H
EARTSTREAM
FR2 is ready
for use. However, in the event that the FR2 has been dropped or mishandled, it is
recommended that the battery be removed and reinstalled to initiate a selftest. If the
FR2 has visible signs of damage, contact Agilent Technologies for technical support.
Battery Insertion Selftest
As described in Chapter 2, Preparing Your H
EARTSTREAM
FR2 for Use, when you
insert the battery in the FR2,
be sure that the defibrillation pads are not connected
to the device. When you insert the battery, a menu is displayed and a two-part
selftest will run unless you make another selection from the menu within 10
seconds. The selftest includes an automatic part and an interactive part.
NOTE:
Under certain circumstances, the behavior of your FR2 will be different. For
example, the menu screen will
not
appear when a battery is inserted if:
•
the defibrillation pads are attached to a patient, indicating that the
H
EARTSTREAM
FR2 is in continued use, or
•
the battery is completely depleted.
The menu screen
will
be displayed, but after 10 seconds the FR2 will go to standby
mode if you make no selection and:
•
less than five minutes have passed since the FR2 was last used, indicating that
the FR2 is still in use.
It is recommended that the full selftest (including the interactive portion) be run
under the following circumstances:
When the FR2 is first put into service and following each use.
Whenever the battery is replaced.
Whenever expired pads are replaced during periodic maintenance.
Whenever the AED may have sustained physical damage.
When you install the battery, the screen tells you whether or not a data card is
installed. If so, a screen message displays how much recording time is left until the
data card is full. (See Chapter 7, Data Management and Review, for how to review the
incident information from the internal memory of the H
EARTSTREAM
FR2 or from a
data card, if one is used.)
RUN SELFTEST
REVIEW INCIDENT
CARD FULL IN XX.XH
GOOD BATTERY
NEXT
IN EMERGENCY
PRESS OFF TO QUIT
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...