Troubleshooting Guide
4-10
Maintaining, Testing, and Troubleshooting Your H
EARTSTREAM
FR2
CHARGES — the total number of full defibrillation charges that have been
provided by this battery, including selftest charges;
BATTERY — a GOOD BATTERY, LOW BATTERY or REPLACE BATTERY
message, as appropriate, and
STATUS — the current status of this battery, displayed in a binary code. Make a
note of this code if technical support is needed.
Troubleshooting Guide
The H
EARTSTREAM
FR2 has several ways of communicating with you if it detects a
problem during periodic selftesting or use. Depending on the nature of the problem,
the FR2 will use audible beeps (“chirps”), voice prompts, screen prompts, and/or its
Status Indicator.
The following tables describe FR2 responses to problems that may occur during use,
and provide recommendations for how to resolve them. Although there is some
overlap in the information presented, for ease of reference the tables separately
address the following types of issues:
Status Indicator Summary — a basic overview of FR2 usability based on the
states shown by the Status Indicator
Status Indicator Details — possible causes of various Status Indicator and other
signals, and recommended actions
Battery and Training & Administration Pack — possible power-related problems
and recommended actions
Defibrillation Pads — possible defibrillation pads problems and recommended
actions
Analysis and Defibrillation — possible problems with rhythm analysis and
defibrillation and recommended actions
It is important to read this information carefully, so you can act quickly in an
emergency.
NOTE:
Perform CPR (if needed) any time there is a delay before the H
EARTSTREAM
FR2 can be used.
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...