M3860A and M3861A H
EARTSTREAM
FR2 Defibrillator
6-1
6
Setup and Advanced Mode Features
Setup Overview
The “setup” of the H
EARTSTREAM
FR2 AED is made up of several programmable
aspects, or parameters, of FR2 operation. Some setup parameters govern specific
features that are not related to the patient care protocol, some are used to define the
automatic patient care protocol used by the FR2, and some provide options for
manual override of the protocol during use.
The FR2 comes with a factory default setup designed to meet the needs of most
users. If desired, your Medical Director can revise the setup. Even if no changes are
made, however, it is a good idea to understand the setup of your FR2 and how the
different parameter settings affect the way the device works.
Non-Protocol Parameters
The parameters listed in the following table enable features of FR2 operation that
are not related to the patient treatment protocol. The table describes each of these
non-protocol parameters, lists the settings available for it, and identifies the default
setting.
PARAMETER
SETTINGS
DEFAULT
DESCRIPTION
SPEAKER
VOLUME
1, 2, 3, 4,
5, 6, 7, 8
8
Sets volume of the FR2’s speaker. 1 is lowest; 8 highest. The
speaker is used for voice prompts and the charge-done
tone.
RECORD VOICE
YES, NO
NO
Enables or disables the audio recording during incident.
Voice recording requires use of a data card.
ECG DISPLAY
(M3860A only)
ON, OFF
ON
Enables (ON) or disables (OFF) ECG display on the screen.
FR2 rhythm analysis does not require ECG display to be on.
(ECG display cannot be changed from default OFF for
M3861A.)
Содержание Heartstream FR2
Страница 1: ...Agilent Heartstream FR2 M3860A M3861A Agilent Technologies www ardusmedical com...
Страница 2: ...User s Guide Agilent M3860A M3861A Heartstream FR2 Semi Automatic External Defibrillator AED...
Страница 9: ...iv Contents Notes...
Страница 17: ...Placing and Securing the HEARTSTREAM FR2 2 6 Preparing Your HEARTSTREAM FR2 for Use Notes...
Страница 23: ...Step 3 Shock Delivery 3 6 Using Your HEARTSTREAM FR2 Notes...
Страница 41: ...Troubleshooting Guide 4 18 Maintaining Testing and Troubleshooting Your HEARTSTREAM FR2 Notes...
Страница 47: ...Safety Considerations 5 6 Clinical and Safety Considerations Notes...
Страница 65: ...Accessories A 2 Notes...
Страница 77: ...Symbols and Controls D 4 Notes...